SICS manual for the SANS I instrument | ||
---|---|---|
<<< Previous | Quick reference Guide | Next >>> |
prints the current positioned attenuator.
positions attenuator <val>. Allowed attenuator numbers are: 0, 1, 2, 3, 4 and 5.
0 : square 50 mm x 50 mm slit, attenuation = 1 |
1 : circular 41 x diameter 0.4 mm slit, attenuation = 1/485 |
2 : circular 9 x diameter 2 mm slit, attenuation = 1/88 |
3 : circular 20 mm diameter slit, attenuation = 1/8 |
4 : circular 30 mm diameter slit, attenuation = 1/3.5 |
5 : circular 15 mm diameter slit, attenuation = 1/?? |
<<< Previous | Home | Next >>> |
Neutron Velocity Selector | Up | Changing collimation |