| Magnet
 | 
|---|
| field at sample | 14.9 Tesla | 
| " without lambda stage | 13.5 Tesla | 
| operating current for 14.9 T | <120 A | 
| maximum allowed vertical magnetic forces | 1000 N | 
| maximum allowed horizontal magnetic forces | 1000 N | 
| homogeneity over a 10 mm diameter sphere | 0.5% | 
| field stability in persistent mode | 0.01 % / hour | 
| Stray Field at 14.9 Tesla
 | 
|---|
| at 1 m horizontal distance | 330 Gauss | 
| at 1 m vertical distance | 660 Gauss | 
| critical distance for loose ferromagnetic parts 500 Gauss/metre
 | 1.40 m vertical 1.20 m horizontal
 | 
| Variable Temperature Insert
 | 
|---|
| temperature range (no field dependend limitations) | 1.6 ... 300 K | 
| temperature stability | +- 0.1 K | 
| automatic needle valve control only above | 3 K | 
| exchange gas needed in sample space (measured at RT) | <15 mbar | 
| Dimensions
 | 
|---|
| sample space diameter above magnet | 49.5 mm | 
| diameter of sample space | 20 mm | 
| max. useable sample height | 20 mm | 
| split access angle | +- 2 deg | 
| sample height for +-2 deg access(visible by neutrons) | 14 mm | 
| horizontal access angle | > 330 deg | 
| distance from base of cryostat to scattering plane
 | 240 mm | 
| total radial aluminium window thickness | 43.4 mm | 
| total radial aluminium in entrance window | 8.6 mm | 
| entrance window size (radiussed corners 3 mm) | 15 x 15 mm | 
| diameter for transfer tube | 12 mm | 
| Cryostat
 | 
|---|
| total weight including cryogens | 650+-30 kg | 
| useable LHe capacity | 55 l | 
| LHe boil off (static mode) | <350 cc/hr | 
| LN2 capacity | 60 l | 
| typical LN2 boil off | 600 cc/hr |