MA15RitaRestrictions

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
MA15

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


This is version 5. It is not the current version, and thus it cannot be edited.
[Back to current version]   [Restore this version]


RITA-2

The operation of MA15 on Rita-II is allowed:

  • between 3.0 and 17.0 meV at a distance of 450 mm or more, up to 14.9 Tesla
  • between 3.0 and 8.0 meV at a distance of 410 mm or more, up to 13.5 Tesla
  • between 8.0 and 17.0 meV at a distance of 410 mm or more, up to 11.5 Tesla

For any other configuration force tests have to be done.

MA15RitaRestrictions/SampleTableDistance.jpg
How the distance has to be measured. In this example the distance is around 400 mm.

TASP

The operation on TASP is allowed up to a field of 8 Tesla for a distance of 122 cm from the sample to the surface of the monochromator shielding, and for A2 angles between -41 and -90.


Attachments:  eigerma15dist.jpg  SampleTableDistance.jpg 
This particular version was published on 15-Dec-2010 09:07:28 UTC by MarkusZolliker.