This is version 5.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
RITA-2
The operation of MA15 on Rita-II is allowed:
- between 3.0 and 17.0 meV at a distance of 450 mm or more, up to 14.9 Tesla
- between 3.0 and 8.0 meV at a distance of 410 mm or more, up to 13.5 Tesla
- between 8.0 and 17.0 meV at a distance of 410 mm or more, up to 11.5 Tesla
For any other configuration force tests have to be done.

How the distance has to be measured.
In this example the distance is around 400 mm.
TASP
The operation on TASP is allowed up to a field of 8 Tesla for a distance of 122 cm from the sample
to the surface of the monochromator shielding, and for A2 angles between -41
and -90.
Attachments:
eigerma15dist.jpg
SampleTableDistance.jpg
This particular version was published on 15-Dec-2010 09:07:28 UTC by MarkusZolliker.
|