MA15RitaRestrictions

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
MA15

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


This is version 3. It is not the current version, and thus it cannot be edited.
[Back to current version]   [Restore this version]


The operation of MA15 on Rita-II is allowed:

  • between 3.0 and 14.7 meV at a distance of 565 mm or more, up to 14.9 Tesla
  • between 3.0 and 8 meV at a distance of 410 mm or more, up to 13.5 Tesla
  • between 3.0 and 5.7 meV at a distance of 410 mm or more, up to 14.9 Tesla

For any other configuration force test have to be done.

MA15RitaRestrictions/SampleTableDistance.jpg
How the distance has to be measured. In this example the distance is around 400 mm.


Attachments:  eigerma15dist.jpg  SampleTableDistance.jpg 
This particular version was published on 26-Jan-2005 10:31:02 UTC by MarkusZolliker.