| MA15 Restrictions on RITA-2
The operation of MA15 on Rita-II is allowed:
 
 
 between 3.0 and 17.0 meV  at a distance of 590 mm or more, up to 14.9 Tesla
 between 3.0 and  8.0 meV  at a distance of 460 mm or more, up to 13.5 Tesla
 between 8.0 and 17.0 meV  at a distance of 460 mm or more, up to 11.5 Tesla
 
For any other configuration force tests have to be done.
 
 How the distance has to be measured.
In this example the distance is around 400 mm.
 
      
 Attachments: 
        
        eigerma15dist.jpg 
        
        SampleTableDistance.jpg
 Printer Friendly 
        Page Info