MA15 Restrictions

InstrumentAllowed Max. Field [Tesla]
RITA-2 11.5 ... 14.9 (see below)
EIGER 10
TASP 8 (see below)

RITA-2

The operation of MA15 on Rita-II is allowed:

  • between 3.0 and 17.0 meV at a distance of 450 mm or more, up to 14.9 Tesla
  • between 3.0 and 8.0 meV at a distance of 410 mm or more, up to 13.5 Tesla
  • between 8.0 and 17.0 meV at a distance of 410 mm or more, up to 11.5 Tesla

For any other configuration force tests have to be done.

MA15RitaRestrictions/SampleTableDistance.jpg
How the distance has to be measured. In this example the distance is around 400 mm.



Attachments:  eigerma15dist.jpg  SampleTableDistance.jpg 
This particular version was published on 20-Apr-2012 09:14:49 UTC by MarkusZolliker.