| This is version 1.  
            It is not the current version, and thus it cannot be edited.[Back to current version]  
            [Restore this version]
 The operation of MA15 on Rita-II is allowed:
 
 
 between 3.0 and 14.7 meV  at a distance of 565 mm or more, up to 14.9 Tesla
 between 3.0 and  8   meV  at a distance of 410 mm or more, up to 13.5 Tesla
 between 3.0 and  5.7 meV  at a distance of 410 mm or more, up to 14.9 Tesla
 
For a other configuration force test have to be done.
 
 How the distance has to be measured.
In this example the distance is around 400 mm. 
 Attachments: 
        
        eigerma15dist.jpg 
        
        SampleTableDistance.jpg
 This particular version was published on 26-Jan-2005 10:29:49 UTC by MarkusZolliker.
 |