This is version 8.
It is not the current version, and thus it cannot be edited.
[Back to current version]
[Restore this version]
Windows
Beam parallel to the field:
full window height | 70 mm |
angular opening on a 13 mm sample* | +/- 15 deg |
angular opening on a 5 mm sample* | +/- 17 deg |
Beam perpendicular to the field:
full window height | 20 mm |
angular opening on a 11 mm sample* | +/- 7.1 deg |
angular opening on a 5 mm sample* | +/- 8 deg |
(*) openings are identical vertical and horizontal,
as the cuts in the coils are cones.
Max. sample diameter | 20 mm |
Sample space diameter at magnet center | 22 mm |
Distance from sample stick cap to magnet center | 1302 mm |
Distance from sample stick cap to bottom of sample space | 1358 mm |
Sample space diameter above coil | 62 mm |
Distance from sample stick cap to end of 62 mm space | 1080 mm (+ 2 mm tolerance) |
Motorized Rotator, distance to add | 159 mm |
Motorized Rotator: from sample stick cap to magnet center | 1461 mm |
Motorized Rotator: from sample stick cap to bottom of sample space | 1517 mm |
Magnetic Field Range
if no other Restriction apply
max Bfield T? | lambdaflow |
9.5 | off |
10.8 | 15 mbar |
This particular version was published on 06-Jul-2011 14:49:58 UTC by MarkusZolliker.
|