horizontal | 200N |
vertical | 500N |
Instrument | field direction parallel to beam | field direction perpendicular to beam | Conditions |
---|---|---|---|
SANS-I | 11 Tesla | 11 Tesla | without spin flipper |
SANS-II | 11 Tesla | 11 Tesla | |
AMOR | 2 Tesla * | * | |
RITA-II | 11 Tesla | 11 Tesla | 45 deg < a2 < 100 deg, d = 450 mm |
On all other Intruments, MA11 can not yet be used.
With new detector gate valve (18.8.2015), at 7.8 Tesla:
horizontal (beam direction) < 50 N, ~ 80 N (perpendicular), 220 N vertical
extrapolate forces at 11 T: 160 N horiz., 440 N vertical
How the distance has to be measured.
In this example the distance is around 400 mm.
Force measurements 14.7.2009 at 8 Tesla: