Above 150 K, the liquid He consumption get very high. For a short time 300 K is possible, but not recommended.
Vertically, the access to the sample is a 30 mm split with 282 mm diameter. This means for example a 10 mm sample can be fully illuminated with a horizontal beam when tilting +- 4 deg.
maximum central field | 1.8T |
max current | 101.6 Amp |
Field/Current | 0.0177 T/A |
homogeneity (sphere 10mm diam) | 2pp100 |
magnet clear bore diam | 35mm |
Split between former halves | 30mm |
Split angles | 0deg |
Switch heater current | 75 mA |
CoilSeparation | 1 Al ring |
Distance field center to cryo base | 19 cm |
field range in T | rate in T/min |
0-1.24 | 0.1063 |
1.24-1.6 | 0.0531 |
1.6-1.8 | 0.0177 |