In our test, we mounted an empty vanadium can in the closed cycle refrigerator CTI2, with a test sensor at the bottom of the container. The 80 K radiation screen was always in place.
temperature difference delta T at 12 K | time delay delta t at 30 K | configuration |
---|---|---|
1 K | 33 min | with helium exchange gas |
0.2 K | 30 min | with helium exchange gas, 12 K screen |
0.2 K | 32 min | with helium exchange gas, 12 K screen and vacuum grease |
2 K | 70 min | without helium exchange gas |
0.5 K | 62 min | without helium exchange gas, but with 12 K screen |