Diff: RITA2SpecialCommandsAndProcedures

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
NewTasSyntax

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 28 and version 7:
Line 3 was replaced by line 3
- By Jari í Hjøllum and Christof Niedermayer, 25-08-06
+ By Christof Niedermayer and Jari í Hjøllum
Line 6 was replaced by line 6
- !!How to log in
+ !!Log in
Removed line 8
-
At line 9 added 1 line.
+
Line 13 was replaced by lines 13-14
- 1) In console write "ssh rita2@rita2.psi.ch" and enter your password
+
+ 1) In console write "ssh rita2@rita2.psi.ch" and enter your password.
At line 15 added 1 line.
+
At line 16 added 1 line.
+
At line 20 added 4 lines.
+
+ !! The SICS Client
+
+
At line 21 added 1 line.
+
Lines 23-26 were replaced by line 31
- Alt#1:
- 1) Go to: http://lns00.psi.ch/
- 2) and choose "The SICS Command Line Client"
- ----
+
Line 28 was replaced by lines 33-34
- 1) In the client click Connect and choose RITA-2 from the list. Alternatly you can choose local or spare if you are located at the RITA-2 instrument computer.
+
+ 1) In the client click Connect and choose RITA-2 from the list.
At line 29 added 2 lines.
+ Alternatly you can choose local or spare if you are located at the RITA-2 instrument computer.
+
At line 30 added 1 line.
+
At line 31 added 1 line.
+
At line 32 added 1 line.
+
At line 36 added 12 lines.
+ !! six
+
+ six is an alternative to the standard Java SICS commandline client. Why an alternative SICS client?
+
+ • It is much faster
+
+ • It does not need the X window system
+
+ • It connects automatically to the instrument if you are logged into the instrument computer
+
+
+ ----
At line 42 added 1 line.
+
Line 46 was replaced by lines 70-71
- Current PSD image can be viewed by sansstatus
+ Current PSD image can be viewed by sansstatus
+
At line 47 added 1 line.
+
At line 55 added 1 line.
+
At line 56 added 1 line.
+
At line 62 added 1 line.
+
At line 66 added 1 line.
+
At line 69 added 2 lines.
+
+
Line 74 was replaced by lines 106-107
- topsistatus displays the content of window 5.
+ Note that topsistatus displays the content of window 5.
+
At line 75 added 1 line.
+
At line 76 added 1 line.
+
At line 77 added 1 line.
+
At line 85 added 1 line.
+
Line 87 was replaced by line 124
- ---===+++===---
+
At line 88 added 1 line.
+
Line 90 was replaced by line 128
- ---===+++===---
+
Line 92 was replaced by lines 130-132
- 1) Make a text file, e.g. ">kate test01.job &"
+
+ 1) Make a text file, e.g. ">kate test01.job &". (Extension can also be test01.tcl)
+
At line 93 added 1 line.
+
Line 95 was replaced by lines 136-151
- 4) In a sics client write : "do test01.job"
+
+ If you do this from the computer next to the experimental area, make sure that you are logged on as user rita2 on rita2.psi.ch and not as sinquser on pc4697. If you have to open a new window just follow the login procedure above.
+
+ 4) In a sics client write : "do test01.job" or "exe test01.tcl"
+
+ ----
+
+ !!Listening to the activity
+
+ To listen to the activity on the spectrometer, open a client and run
+
+ > config listen 1
+
+ To turn it off again, type
+
+ > config listen 0
At line 96 added 9 lines.
+ Of course this command can also be run on a already open client.
+
+
+
+ > commandlog tail x
+
+ shows the last x sics message lines up to a maximum of 999 lines, allowing you to check for errors from any sics client.
+
+
At line 103 added 8 lines.
+
+ To synchronize the instrument parameter execute the command sync
+
+
+ Alternatively use
+
+ six -s
+
At line 105 added 46 lines.
+
+ ----
+
+ !!2-axis and 3-axis mode
+
+ The spectrometer can be run in two modes, 2-axis and triple-axis.
+
+ There are 2 triple-axis modes: ki and kf. Usually the kf option is used.
+
+ There is one 2-axis mode: elastic.
+
+ To read which mode the spectrometer is in write
+
+ > tasub const
+
+ To change the mode write for instance
+
+ > tasub const elastic
+
+ ----
+
+
+
+ !!Lambda/2 mode
+
+ To use Lambda /2 neutrons without changing A2 just divide the monochromator d-spacing by a factor of 2, i.e. go from 3.354 A to 1.677 A
+
+ > tasub mono dd 1.677
+
+ > tasub update
+
+
+ Next drive Ei and Ef to 4 times the initial energy.
+ For example for 5 meV = 4.04 A drive Ei and Ef to 20 meV =2.02 A
+
+ Note, that the primary instrument, i.e. the monochromator will not move. The analyzer however will move, so make sure that any field from a magnet is set to zero.
+
+
+ To go back to 5 meV
+
+ > tasub mono dd 3.354
+
+ > tasub update
+
+
+
At line 107 added 1 line.
+
Lines 116-117 were replaced by lines 236-238
- For PSI motors and encoded RISOE motors (2t and 2ta) repositioning should be done by the command sp, which changes the softzero value.
- sp motorname value, eg. sp omm 37.07
+ For PSI motors (monochromator motors omm (a1), 2tm(a2), cum)
+ and encoded RISOE motors (2t(a4) and 2ta)
+ repositioning should be done by the command sp, which changes the softzero value.
At line 118 added 1 line.
+ sp motorname value, eg. sp omm 37.07
Removed lines 120-122
- Repositioning of unencoded RISOE motors can be done by the command
- motorname putpos value, e.g. oma putpos 74.17
- (This is however at the present state of development not working properly for all motors)
At line 125 added 1 line.
+
At line 128 added 1 line.
+
At line 129 added 1 line.
+
At line 137 added 1 line.
+
At line 138 added 1 line.
+
Line 146 was replaced by line 270
- Caution !!!!!!!!!!!
+ !Caution !!!!!!!!!!!
At line 149 added 1 line.
+
At line 153 added 1 line.
+ -----
Line 155 was replaced by lines 281-324
- ----
+
+ !! Disabling and enabling motors
+
+ For all the motors from the DeltaTau controllers, to enable them in SIX you just type, for example
+
+ sgu enable 1
+
+ and to inhibit them, type e.g.
+
+ sgu enable 0
+
+ To list which (of the DeltaTau) motors are enabled or inhibited, just type
+
+ enable
+
+
+ To disable or enable the sample table, i.e. the morors tu,tl,sgu and sgl you can use the commands disabletable and enabletable
+
+ To disbale or enable the slits use disbaleslits or enableslits.
+ You have to disable the motors, before you physically disconnect them, i,e. when removing the sample table or the slits in case of cryomagnet operation.
+
+
+ If you enable or inhibit the motor manually using the hardware switch, it may take up to 30 seconds for SIX to update this parameter and to realise what you've done.
+
+ ML has also stopped SIX from automatically enabling all the motors when SIX is started.
+
+ -----
+
+ !!Filters
+
+ Be and BeO after the sample and before the analyzer tank filters can be moved in and out of the beam by the following commands:
+
+ bef in
+
+ bef out
+
+ The rotation of the filters is controlled by the command brot.
+
+ -----
+
+
+ !!Threshold for PSI counter box
+
+ The PSI counterbox el737 is accessible from sics.
At line 156 added 1 line.
+ The following commands are available.
At line 157 added 18 lines.
+
+ el737 threshold
+
+ reads the current threshold
+
+ el737 threshold value
+
+ sets a new threshold.
+
+ Typical values are 20 to 100. These numbers roughly correspond to the proton current on SINQ.
+
+ This threshold prevents collecting background data in case of no beam or a very low beam.
+
+ The analog in TASCOM was the command CIBO and manual setting of the box.
+
+ -----
+
+
At line 162 added 1 line.
+
At line 166 added 1 line.
+
At line 167 added 1 line.
+
At line 170 added 6 lines.
+ In case of problems with the communication between sics and sea you can try the following.
+
+ In sics use the command connect_sea.
+
+ In sea try the command connect_sics
+
Lines 172-173 were replaced by line 369
- A summary of the commands is foud at
- http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=NewTasSyntax
+ -----
At line 174 added 4 lines.
+ A summary of the commands is found at
+
+ [http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=NewTasSyntax]
+
Line 179 was replaced by line 379
-
+ ----
Line 185 was replaced by lines 385-386
- http://users.belgacom.net/bruno.champagne/tcl.html
+
+ [http://users.belgacom.net/bruno.champagne/tcl.html]
At line 190 added 1 line.
+
At line 191 added 1 line.
+
At line 192 added 1 line.
+
At line 196 added 1 line.
+
At line 199 added 1 line.
+
At line 206 added 1 line.
+
At line 207 added 1 line.
+
Removed lines 211-218
- monit status sicsserver/all/somethingelse
- monit stop sicsserver/all/somethingelse
- monit start sicsserver/all/somethingelse
- monit quit
- monit -h
-
-
- ----
At line 219 added 1 line.
+ monit status sicsserver/all/somethingelse
Line 221 was replaced by line 422
- !!Switching to TASCOM (hopefully not needed)
+ monit stop sicsserver/all/somethingelse
At line 222 added 1 line.
+ monit start sicsserver/all/somethingelse
Line 224 was replaced by line 426
- If you want to switch to TASCOM, stop the Sicsserver
+ monit restart sicsserver
Removed line 226
- monit stop all
Line 229 was replaced by line 430
- Log into rita2 as user tascom and enter your password [*********]
+ monit -h
Removed lines 231-232
- startserial (starts necessary portserver)
- startecs (starts tecs for temperature control)
Lines 235-236 were replaced by lines 434-435
- If you want to switch back to SICS quit tascom and
- kill the portserver
+ ----
+ ----
Removed line 238
- Look for process Serportserv, notice its PID and execute:
Removed line 240
- kill -9 PID

Back to RITA2SpecialCommandsAndProcedures, or to the Page History.