At line 0 added 1 line.
+ !!! Quick helper for RITA-2 and SICS.
Line 2 was replaced by line 3
- Quick helper for RITA-2 and SICS.
+ By Christof Niedermayer and Jari í Hjøllum
Removed line 4
- By Jari í Hjøllum and Christof Niedermayer, 25-08-06
At line 5 added 1 line.
+ !!Log in
Removed lines 7-9
- How to log in
-
-
Removed line 11
- 1) Log in as user: rita2 [********]
At line 12 added 1 line.
+ 1) Log in as user: rita2 (password)
Lines 15-16 were replaced by lines 13-15
- 1) In console write "ssh rita2@rita2.psi.ch" and enter your password [*******]
- ---===+++===---
+
+ 1) In console write "ssh rita2@rita2.psi.ch" and enter your password.
+
At line 17 added 1 line.
+
At line 18 added 1 line.
+
Removed lines 20-21
- ---===+++===---
- ---===+++===---
At line 22 added 1 line.
+ ----
At line 23 added 4 lines.
+
+ !! The SICS Client
+
+
At line 24 added 1 line.
+
Lines 26-29 were replaced by line 31
- Alt#1:
- 1) Go to: http://lns00.psi.ch/
- 2) and choose "The SICS Command Line Client"
- ---===+++===---
+
Lines 31-32 were replaced by lines 33-37
- 1) In the client click Connect and choose RITA-2 from the list. Alternatly you can choose local or spare if you are located at the RITA-2 instrument computer.
- ---===+++===---
+
+ 1) In the client click Connect and choose RITA-2 from the list.
+
+ Alternatly you can choose local or spare if you are located at the RITA-2 instrument computer.
+
At line 33 added 1 line.
+
Lines 35-36 were replaced by lines 41-48
- a) rita2user [*******]
- b) lnsmanager [******]
+
+ a) rita2user (password)
+
+ b) lnsmanager (password)
+
+
+ ----
+ !! six
At line 37 added 1 line.
+ six is an alternative to the standard Java SICS commandline client. Why an alternative SICS client?
At line 38 added 1 line.
+ • It is much faster
Line 40 was replaced by line 54
- Viewing and plotting data
+ • It does not need the X window system
At line 41 added 7 lines.
+ • It connects automatically to the instrument if you are logged into the instrument computer
+
+
+ ----
+
+ !!Viewing and plotting data
+
At line 44 added 1 line.
+
Line 48 was replaced by lines 70-71
- Current PSD image can be viewed by sansstatus
+ Current PSD image can be viewed by sansstatus
+
At line 49 added 1 line.
+
At line 57 added 1 line.
+
At line 58 added 1 line.
+
At line 64 added 1 line.
+
At line 68 added 1 line.
+
Line 74 was replaced by lines 102-106
- Windows.
+ ----
+
+ !!Windows
+
+ Note that topsistatus displays the content of window 5.
Removed line 76
- topsistatus displays the content of window 5.
At line 77 added 1 line.
+
At line 78 added 1 line.
+
Removed line 80
- window5 60:85:30:100
At line 81 added 1 line.
+ window5 60:85:30:100
At line 83 added 1 line.
+ ----
Line 85 was replaced by line 119
- Batch files
+ !!Batch files
At line 87 added 1 line.
+
Line 89 was replaced by line 124
- ---===+++===---
+
At line 90 added 1 line.
+
Line 92 was replaced by line 128
- ---===+++===---
+
Line 94 was replaced by lines 130-132
- 1) Make a text file, e.g. ">kate test01.job &"
+
+ 1) Make a text file, e.g. ">kate test01.job &". (Extension can also be test01.tcl)
+
At line 95 added 1 line.
+
Line 97 was replaced by lines 136-147
- 4) In a sics client write : "do test01.job"
+
+ If you do this from the computer next to the experimental area, make sure that you are logged on as user rita2 on rita2.psi.ch and not as sinquser on pc4697. If you have to open a new window just follow the login procedure above.
+
+ 4) In a sics client write : "do test01.job" or "exe test01.tcl"
+
+ ----
+
+ !!Listening to the activity
+
+ To listen to the activity on the spectrometer, open a client and run
+
+ > config listen 1
At line 98 added 1 line.
+ To turn it off again, type
At line 99 added 1 line.
+ > config listen 0
Line 101 was replaced by line 153
- Simulation Mode
+ Of course this command can also be run on a already open client.
At line 102 added 11 lines.
+
+
+ > commandlog tail x
+
+ shows the last x sics message lines up to a maximum of 999 lines, allowing you to check for errors from any sics client.
+
+
+ ----
+
+ !!Simulation Mode
+
At line 105 added 8 lines.
+
+ To synchronize the instrument parameter execute the command sync
+
+
+ Alternatively use
+
+ six -s
+
At line 108 added 33 lines.
+ ----
+
+ !!2-axis and 3-axis mode
+
+ The spectrometer can be run in two modes, 2-axis and triple-axis.
+
+ There are 2 triple-axis modes: ki and kf. Usually the kf option is used.
+
+ There is one 2-axis mode: elastic.
+
+ To read which mode the spectrometer is in write
+
+ > tasub const
+
+ To change the mode write for instance
+
+ > tasub const elastic
+
+ ----
+
+
+
+ !!Lambda/2 mode
+
+ To use Lambda /2 neutrons without changing A2 just divide the monochromator d-spacing by a factor of 2, i.e. go from 3.354 A to 1.677 A
+
+ > tasub mono dd 1.677
+
+ > tasub update
+
+
+ Next drive Ei and Ef to 4 times the initial energy.
+ For example for 5 meV = 4.04 A drive Ei and Ef to 20 meV =2.02 A
Line 110 was replaced by line 214
- HDF Viewer
+ Note, that the primary instrument, i.e. the monochromator will not move. The analyzer however will move, so make sure that any field from a magnet is set to zero.
At line 111 added 14 lines.
+
+ To go back to 5 meV
+
+ > tasub mono dd 3.354
+
+ > tasub update
+
+
+
+ ----
+
+
+ !!HDF Viewer
+
At line 113 added 1 line.
+ ----
At line 114 added 1 line.
+ !!"Remoing"
Line 116 was replaced by lines 236-238
- "Remoing"
+ For PSI motors (monochromator motors omm (a1), 2tm(a2), cum)
+ and encoded RISOE motors (2t(a4) and 2ta)
+ repositioning should be done by the command sp, which changes the softzero value.
Removed line 118
- For PSI motors and encoded RISOE motors (2t and 2ta) repositioning should be done by the command sp, which changes the softzero value.
Removed lines 122-125
- Repositioning of unencoded RISOE motors can be done by the command
- motorname putpos value, e.g. oma putpos 74.17
- (This is however at the present state of development not working properly for all motors)
-
At line 127 added 1 line.
+
At line 130 added 1 line.
+
Removed line 132
- ca5 softlowerlim -200
At line 133 added 1 line.
+ ca5 softlowerlim -200
At line 135 added 1 line.
+ ----
Line 137 was replaced by line 257
- RITA modes
+ !!RITA modes
At line 139 added 1 line.
+
At line 140 added 1 line.
+
Line 148 was replaced by line 270
- Caution !!!!!!!!!!!
+ !Caution !!!!!!!!!!!
At line 151 added 1 line.
+
At line 155 added 44 lines.
+ -----
+
+
+ !! Disabling and enabling motors
+
+ For all the motors from the DeltaTau controllers, to enable them in SIX you just type, for example
+
+ sgu enable 1
+
+ and to inhibit them, type e.g.
+
+ sgu enable 0
+
+ To list which (of the DeltaTau) motors are enabled or inhibited, just type
+
+ enable
+
+
+ To disable or enable the sample table, i.e. the morors tu,tl,sgu and sgl you can use the commands disabletable and enabletable
+
+ To disbale or enable the slits use disbaleslits or enableslits.
+ You have to disable the motors, before you physically disconnect them, i,e. when removing the sample table or the slits in case of cryomagnet operation.
+
+
+ If you enable or inhibit the motor manually using the hardware switch, it may take up to 30 seconds for SIX to update this parameter and to realise what you've done.
+
+ ML has also stopped SIX from automatically enabling all the motors when SIX is started.
+
+ -----
+
+ !!Filters
+
+ Be and BeO after the sample and before the analyzer tank filters can be moved in and out of the beam by the following commands:
+
+ bef in
+
+ bef out
+
+ The rotation of the filters is controlled by the command brot.
+
+ -----
+
+
+ !!Threshold for PSI counter box
At line 156 added 1 line.
+ The PSI counterbox el737 is accessible from sics.
At line 157 added 1 line.
+ The following commands are available.
Line 160 was replaced by line 329
- Temperature control
+ el737 threshold
At line 161 added 17 lines.
+ reads the current threshold
+
+ el737 threshold value
+
+ sets a new threshold.
+
+ Typical values are 20 to 100. These numbers roughly correspond to the proton current on SINQ.
+
+ This threshold prevents collecting background data in case of no beam or a very low beam.
+
+ The analog in TASCOM was the command CIBO and manual setting of the box.
+
+ -----
+
+
+ !!Temperature control
+
At line 164 added 1 line.
+
At line 168 added 1 line.
+
At line 169 added 1 line.
+
At line 172 added 1 line.
+ In case of problems with the communication between sics and sea you can try the following.
Lines 174-175 were replaced by line 364
- A summary of the commands is foud at
- http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=NewTasSyntax
+ In sics use the command connect_sea.
Line 177 was replaced by lines 366-369
- There you also find a detailled discription on how to align a single crystal (chapter UB matrix stuff).
+ In sea try the command connect_sics
+
+
+ -----
At line 178 added 1 line.
+ A summary of the commands is found at
At line 179 added 1 line.
+ [http://lns00.psi.ch/sinqwiki/Wiki.jsp?page=NewTasSyntax]
At line 180 added 2 lines.
+ There you also find a detailled discription on how to align a single crystal (chapter UB matrix stuff).
+
At line 181 added 2 lines.
+ ----
+ ----
Removed lines 183-185
- ...___---===+++===---___...
- | Advanced stuff |
- ...___---===+++===---___...
At line 186 added 1 line.
+ !! Advanced stuff
Removed line 188
- ---===+++===---
Line 190 was replaced by lines 385-386
- http://users.belgacom.net/bruno.champagne/tcl.html
+
+ [http://users.belgacom.net/bruno.champagne/tcl.html]
Line 192 was replaced by line 388
- ---===+++===---
+ ----
Line 196 was replaced by lines 392-394
- 1) Log in as user: rita2 [********]
+
+ 1) Log in as user: rita2 (password)
+
At line 197 added 1 line.
+
Line 199 was replaced by lines 398-399
- ---===+++===---
+
+
At line 200 added 1 line.
+
Line 202 was replaced by line 403
- ---===+++===---
+
At line 203 added 1 line.
+
At line 210 added 1 line.
+
At line 211 added 1 line.
+
Removed lines 215-221
- monit status sicsserver/all/somethingelse
- monit stop sicsserver/all/somethingelse
- monit start sicsserver/all/somethingelse
- monit quit
- monit -h
-
-
Line 223 was replaced by line 420
-
+ monit status sicsserver/all/somethingelse
Line 225 was replaced by line 422
- Switching to TASCOM (hopefully not needed)
+ monit stop sicsserver/all/somethingelse
At line 226 added 1 line.
+ monit start sicsserver/all/somethingelse
Line 228 was replaced by line 426
- If you want to switch to TASCOM, stop the Sicsserver
+ monit restart sicsserver
Removed line 230
- monit stop all
Line 233 was replaced by line 430
- Log into rita2 as user tascom and enter your password [*********]
+ monit -h
Removed lines 235-236
- startserial (starts necessary portserver)
- startecs (starts tecs for temperature control)
Lines 239-240 were replaced by lines 434-435
- If you want to switch back to SICS quit tascom and
- kill the portserver
+ ----
+ ----
Removed line 242
- Look for process Serportserv, notice its PID and execute:
Removed line 244
- kill -9 PID