At line 199 added 28 lines.
+
+
+ !!Lambda/2 mode
+
+ To use Lambda /2 neutrons without changing A2 just divide the monochromator d-spacing by a factor of 2, i.e. go from 3.354 A to 1.677 A
+
+ > tasub mono dd 1.677
+
+ > tasub update
+
+
+ Next drive Ei and Ef to 4 times the initial energy.
+ For example for 5 meV = 4.04 A drive Ei and Ef to 20 meV =2.02 A
+
+ Note, that the primary instrument, i.e. the monochromator will not move. The analyzer however will move, so make sure that any field from a magnet is set to zero.
+
+
+ To go back to 5 meV
+
+ > tasub mono dd 3.354
+
+ > tasub update
+
+
+
+ ----
+
+