Diff: MA7Restrictions

Search SINQ Wiki:
SINQ LIN

SINQ Wiki
- Main Page
- Search SINQ Wiki
- Sample Environment
- Probenumg. Intern
- Troubleshooting SICS

This Page
- Page Info
- Printer Friendly

Referenced by
MA7

Wiki Info
- Unused pages
- Undefined pages
- RecentChanges
- Page Index
- System Info
- JSPWiki Docu
- SandBox
- OneMinuteWiki
- Create a New Page




JSPWiki v2.0.52


Difference between version 53 and version 33:
Line 10 was replaced by lines 10-11
- |AMOR |4
+ |AMOR |4 | field perpendicular beam
+ |AMOR |6.8 | field parallel beam
Lines 14-15 were replaced by lines 15-18
- |TASP |5.5 | see below
- |EIGER |6.8 / 6.2 (full field)|
+ |TASP |6.8 | see below (A3 does not move > 4 Tesla)
+ |EIGER |6.8 / 6.2 (full field)|goniometer removed
+ |ZEBRA |6.8 / 6.2 (full field)|L1 = 1500 mm, L2 = 785 mm (1D detector)
+ |ZEBRA | 5 / 5 |L1 = 1000 mm, L2 = 785 mm (1D detector)
Lines 40-75 were replaced by line 43
- Provisional force measuremnents on AMOR have shown that
- the operation of MA7 on AMOR is allowed with a field perpendicular
- to the beam at least __up to 4 Tesla__.
- The distance to the shielding wall between AMOR and SANS-II has
- to be at least 540 mm, measured from the aluminium window
- (or about 870 mm to the magnet centre).
-
- ''Torque Calculations''
-
- In general, the forces are not only radial, but also tangential,
- inducing a torque onto the magnet. For estimating the forces
- on the fixation, we need to take into account the distance of
- the fixation points from the coil center.
-
- For MA7, with a mean coil radius of about 100 m and fixation points
- at about 150 mm for the coil center, we get the following:
-
- Ro: object distance (mm)\\
- phi: angle (deg) between field direction and direction of object, for maximum force on fixation\\
- Fp: force when object on magnet axis (in field direction)
- Ft: maximum total force on fixation
-
- ||Ro ||phi ||Ft / Fp
- |<600 | 0 | 1.0
- | 800 | 20 | 1.04
- | 1000 | 27 | 1.13
- | 1200 | 32 | 1.25
-
- For the distance to the shielding wall we have to add 1/5 of the edge
- distance for the mean distance of the material that has to be taken
- into account (870 * 6/5 = 1044 mm).
-
- Measured force at 3.4 Tesla: 180 N.\\
- Extrapolated to 5 Tesla: 390 N.\\
- Factor 1.2 for including torque: 470 N.\\
- With a reasonable security margin, we allow to go __up to 4 Tesla__.
+ Force tests after major instrument upgrade 2023-11-21:
At line 76 added 4 lines.
+ Allowed field when perpendicular to beam: __6.8 Tesla__ (forces 300 N horizontally, 480 N vertically)
+
+ Field parallel to field: __3 Tesla__ (horizontal force 280 N, guide field is moving, not secured enough! this should be fixed)
+
Lines 80-84 were replaced by lines 52-53
- Distance: sample --> analyzer is set to max value\\
- This will give Be-filter ~22 cm from magnet surface (see photo)\\
- \\
- Distance: sample table --> monochromator ~ 57 cm (see photo)
-
+ Beryllium Filter Distance 27 cm (see photo)\\
+ Sample table - Monochromator - 73 cm\\
Lines 86-88 were replaced by lines 55-62
- |-92 < A2 < -41 (Ei=14.7 meV)|<= 90? | __5.5 Tesla__
- |-92 < A2 < -56 (Ei=8 meV) |<= 112 | __5.5 Tesla__
- |-92 < A2 < -80 (Ei=3.5 meV) |<= 119 | __5.5 Tesla__
+ |-92 < A2 < -41 (Ei=14.7 meV)|<= 119 | __6.8 Tesla__
+
+ __WARNING: Motor A3 may not move with a field > 4 Tesla__
+
+ This data is derived from the force test 7. July 2015:\\
+ 4.8 Tesla, horizontal force 150 N (towards Beryllium filter), vertical force 180 N.
+
+ [MA7Restrictions/taspFilterMA7.png][MA7Restrictions/distanceTASP.jpg]

Back to MA7Restrictions, or to the Page History.