Line 10 was replaced by lines 10-11 |
- |AMOR |4 |
+ |AMOR |4 | field perpendicular beam |
+ |AMOR |6.8 | field parallel beam |
Lines 13-14 were replaced by lines 14-18 |
- |TASP |3.3 ... 5 |see below |
- |EIGER |6.8 / 6.2 (full field)| |
+ |SANS-II |6.8 / 6.2 (full field)| |
+ |TASP |6.8 | see below (A3 does not move > 4 Tesla) |
+ |EIGER |6.8 / 6.2 (full field)|goniometer removed |
+ |ZEBRA |6.8 / 6.2 (full field)|L1 = 1500 mm, L2 = 785 mm (1D detector) |
+ |ZEBRA | 5 / 5 |L1 = 1000 mm, L2 = 785 mm (1D detector) |
At line 17 added 1 line. |
+ Goniometer removed.\\ |
Lines 38-73 were replaced by line 43 |
- Provisional force measuremnents on AMOR have shown that |
- the operation of MA7 on AMOR is allowed with a field perpendicular |
- to the beam at least __up to 4 Tesla__. |
- The distance to the shielding wall between AMOR and SANS-II has |
- to be at least 540 mm, measured from the aluminium window |
- (or about 870 mm to the magnet centre). |
- |
- ''Torque Calculations'' |
- |
- In general, the forces are not only radial, but also tangential, |
- inducing a torque onto the magnet. For estimating the forces |
- on the fixation, we need to take into account the distance of |
- the fixation points from the coil center. |
- |
- For MA7, with a mean coil radius of about 100 m and fixation points |
- at about 150 mm for the coil center, we get the following: |
- |
- Ro: object distance (mm)\\ |
- phi: angle (deg) between field direction and direction of object, for maximum force on fixation\\ |
- Fp: force when object on magnet axis (in field direction) |
- Ft: maximum total force on fixation |
- |
- ||Ro ||phi ||Ft / Fp |
- |<600 | 0 | 1.0 |
- | 800 | 20 | 1.04 |
- | 1000 | 27 | 1.13 |
- | 1200 | 32 | 1.25 |
- |
- For the distance to the shielding wall we have to add 1/5 of the edge |
- distance for the mean distance of the material that has to be taken |
- into account (870 * 6/5 = 1044 mm). |
- |
- Measured force at 3.4 Tesla: 180 N.\\ |
- Extrapolated to 5 Tesla: 390 N.\\ |
- Factor 1.2 for including torque: 470 N.\\ |
- With a reasonable security margin, we allow to go __up to 4 Tesla__. |
+ Force tests after major instrument upgrade 2023-11-21: |
At line 74 added 4 lines. |
+ Allowed field when perpendicular to beam: __6.8 Tesla__ (forces 300 N horizontally, 480 N vertically) |
+ |
+ Field parallel to field: __3 Tesla__ (horizontal force 280 N, guide field is moving, not secured enough! this should be fixed) |
+ |
Lines 77-78 were replaced by lines 51-57 |
- Distance: sample --> analyzer is set to max value\\ |
- Be-filter installed ~22 cm from magnet surface (see photo)\\ |
+ Goniometer removed.\\ |
+ Beryllium Filter Distance 27 cm (see photo)\\ |
+ Sample table - Monochromator - 73 cm\\ |
+ ||A2 || a4 || Allowed field |
+ |-92 < A2 < -41 (Ei=14.7 meV)|<= 119 | __6.8 Tesla__ |
+ |
+ __WARNING: Motor A3 may not move with a field > 4 Tesla__ |
Lines 80-83 were replaced by lines 59-60 |
- ||A2 || distance sample table --> \\monochromator (see photo)|| Allowed field |
- |-92 < A2 < -41 (Ei=14.7 meV)|>=57 cm | __5.5 Tesla__ |
- |-92 < A2 < -56 (Ei=8 meV)|>=57 cm | __5.5 Tesla__ |
- |-92 < A2 < -80 (Ei=3.5 meV)|>=57 cm | __5.5 Tesla__ |
+ This data is derived from the force test 7. July 2015:\\ |
+ 4.8 Tesla, horizontal force 150 N (towards Beryllium filter), vertical force 180 N. |
At line 84 added 2 lines. |
+ [MA7Restrictions/taspFilterMA7.png][MA7Restrictions/distanceTASP.jpg] |
+ |
At line 95 added 4 lines. |
+ |
+ !Force Test on SANS-II 21.5.2015 |
+ |4.9 T perpendicular to beam|300 N vertical|140 N horizontal |
+ |4.9 T parallel to beam|300 N vertical|60 N horizontal |