Line 1 was replaced by line 1 |
- !RITA-2 |
+ !MA15 Restrictions on RITA-2 |
Lines 5-7 were replaced by lines 5-7 |
- * between 3.0 and 17.0 meV at a distance of 450 mm or more, up to 14.9 Tesla |
- * between 3.0 and 8.0 meV at a distance of 410 mm or more, up to 13.5 Tesla |
- * between 8.0 and 17.0 meV at a distance of 410 mm or more, up to 11.5 Tesla |
+ * between 3.0 and 17.0 meV at a distance of 590 mm or more, up to 14.9 Tesla |
+ * between 3.0 and 8.0 meV at a distance of 460 mm or more, up to 13.5 Tesla |
+ * between 8.0 and 17.0 meV at a distance of 460 mm or more, up to 11.5 Tesla |
Removed lines 15-23 |
- !TASP |
- |
- The operation on TASP is allowed __up to a field of 8 Tesla__ for a distance of 122 cm from the sample |
- to the surface of the monochromator shielding, and for A2 angles between -41 |
- and -90. |
- |
- !EIGER |
- |
- Operation permited to 10T at 270mm distance between sampletable and Monochromator bolck mounting. |