| Line 1 was replaced by line 1 | 
| - !!MA15 Restrictions | 
| + !MA15 Restrictions on RITA-2 | 
| Removed lines 3-9 | 
| - ||Instrument||Allowed Max. Field [[Tesla] | 
| - |RITA-2 |11.5 ... 14.9 (see below) | 
| - |EIGER  |10 | 
| - |TASP   | 8 (see below) | 
| - | 
| - !RITA-2 | 
| - | 
| Lines 12-14 were replaced by lines 5-7 | 
| - * between 3.0 and 17.0 meV  at a distance of 450 mm or more, up to 14.9 Tesla | 
| - * between 3.0 and  8.0 meV  at a distance of 410 mm or more, up to 13.5 Tesla | 
| - * between 8.0 and 17.0 meV  at a distance of 410 mm or more, up to 11.5 Tesla | 
| + * between 3.0 and 17.0 meV  at a distance of 590 mm or more, up to 14.9 Tesla | 
| + * between 3.0 and  8.0 meV  at a distance of 460 mm or more, up to 13.5 Tesla | 
| + * between 8.0 and 17.0 meV  at a distance of 460 mm or more, up to 11.5 Tesla | 
| Removed lines 22-29 | 
| - !TASP | 
| - | 
| - The operation on TASP is allowed __up to a field of 12 Tesla__ for A2 angles between -45 and -100. | 
| - The distance to the monochromator must be 630 mm or more, and the distance to the beryllium filter | 
| - 250 mm. See pictures how to measure: | 
| - [MA15TASPRestrictions] | 
| - | 
| - |