Line 1 was replaced by line 1 |
- !!MA15 Restrictions |
+ !MA15 Restrictions on RITA-2 |
Removed lines 3-9 |
- ||Instrument||Allowed Max. Field [[Tesla] |
- |RITA-2 |11.5 ... 14.9 (see below) |
- |EIGER |10 |
- |TASP | 8 (see below) |
- |
- !RITA-2 |
- |
Lines 12-14 were replaced by lines 5-7 |
- * between 3.0 and 17.0 meV at a distance of 450 mm or more, up to 14.9 Tesla |
- * between 3.0 and 8.0 meV at a distance of 410 mm or more, up to 13.5 Tesla |
- * between 8.0 and 17.0 meV at a distance of 410 mm or more, up to 11.5 Tesla |
+ * between 3.0 and 17.0 meV at a distance of 590 mm or more, up to 14.9 Tesla |
+ * between 3.0 and 8.0 meV at a distance of 460 mm or more, up to 13.5 Tesla |
+ * between 8.0 and 17.0 meV at a distance of 460 mm or more, up to 11.5 Tesla |
Removed lines 22-33 |
- !TASP |
- |
- The operation on TASP is allowed __up to a field of 12 Tesla__ for A2 angles between -45 and -100. |
- The distance to the monochromator must be 630 mm or more, and the distance to the beryllium filter |
- 250 mm. See pictures how to measure: |
- [MA15TASPRestrictions] |
- |
- !EIGER |
- |
- Operation permitted to 10T at 270mm distance between sample table and monochromator block mounting. |
- This limit might be increased after exchangeing some magnetic parts in the sample table.\\ |
- [eigerma15dist.jpg] |