DMC and HRPT specific commands and devices


DMC and HRPT specific commands

StoreData
Does what it says. Writes the current state of the instrument including counts to a NeXus data file.
count mode preset
Does a count operation in mode with a preset of preset. The parameters are optional. If they are not given the count will be started with the current setting in the histogram memory object. After the count, StoreData will be automatically called.
Repeat num mode preset.
Calls count num times. num is a required parameter. The other two are optional and are handled as described above for count.
scan motor start step n mode preset
This command allows for scanning a motor against monitors 0 and 1. This command may only be used by managers. Its only purpose is to facilitate the adjustment of the instrument. In order to obtain a copy of the scan results, a user must take great care to enable command logging. The parameters are: a motor to be scanned, a start value for the scan, a step width for the scan, the number of scan points, optionally a count mode and a preset value for counting. Both these parameters have meanings as described above for the count command.

DMC motor list

OmegaM, A1
Omega monochromator.
TwoThetaM, A2
Two Theta monochromator
MonoX
X-translation table of the monochromator.
MonoY
Y-translation table for the monochromator.
CurveM
Monochromator curvature.
MonoPhi
Phi angle of the monochromator.
MonoChi
Chi angle for the monochromator.
Table, A3
Sample rotation.
TwoThetaD, A4
Two Theta detector.

Other DMC and HRPT devices

banana
Histogram memory.
counter
EL737 counter box.

DMC and HRPT Variables

Instrument
Instrument name.
Title
Experiment title.
comment1, comment2, comment3
comment lines to be stored with the data.
Collimation
Text line describing collimators in use.
User
User name.
Adress
User adress
phone
User phone number.
fax
User fax number
email
User email adress.
Sample
Sample name
sample_mur
Absorption coefficient of sample.